Jandel RM3-AR Test Unit with PC Software
Combined constant current source and digital voltmeter designed specifically for the four point probe measurement
The RM3-AR Test Unit is a specialty electronics instruments designed specifically for the four point probe measurement. It features high accuracy, an excellent range, and many features which simplify the four point probing measurement. The following are features of the RM3-AR Test Unit:
- The measurement range of the RM3-AR Test Unit is from 1 milliohm-per-square (10^-3) up to 5 x 10^8 ohms-per-square with 0.3% accuracy. The volume resistivity range is from 1 milliohm-cm (10^-3) up to 10^6 ohms-cm.
- The RM3-AR includes PC control software which can be used for data logging (storing data in the CSV format) and measurement conversion to ohms-per-square or ohms-cm.
- The RM3-AR reads-out directly in ohms-per-square (or toggle to millivolts) without requiring the use of the software or a PC.
- The RM3-AR has onboard non-volatile memory so that up to 50 measurements can be stored internally and then downloaded and saved all at one time using the software. Alternately, each measurement can be saved to a PC as it is made.
- The RM3-AR has an auto-range button that can be used to automatically determine the optimum input current for a given material without using the trial and error method.
- The RM3-AR has forward (FWD) and reverse (REV) buttons to reverse the direction of current flow. A common way to determine if a measurement is valid is to reverse the direction of current flow and then check to see if the forward and reverse voltage readings correlate well, i.e., the values should be similar, but with the reverse current voltage being a negative value.
|
 (click image to see larger photo)
 Shown here displaying result in ohms-per-square
|
SPECIFICATIONS:
- Superior Current Source
- 10nA to 100mA (99.999mA) current source selectable in steps to 5 decimal place resolution
- Current set numeric keypad
- 4 default preset current programs (user programmable)
Superior Inbuilt DVM
- Input Impedance 1,000,000,000,000 ohms
- Input Bias current 4pA
- DVM 1300mV range and 130mV range
- 130mV accuracy
- 0.2% +/- 5uV resolution (10uV or 1uV) range
- 1300mV accuracy 0.2%+/-100uV resolution
- 100uV Ohms/Square
- Rapid Zeroing null function for DVM
Download a one page product brochure (83K pdf file) regarding the RM3-AR Test Unit
Download the instruction manual for the RM3-AR which includes software instructions (700k pdf file)
|
Features
- 28 Key high quality Keypad
- 16x2 line LCD Display for simultaneous indication of Set Current and either Ohms/Sq or mV
- Intuitive operation
- Microprocessor controlled
- Reduced Footprint
- Robust Attractive ABS Case
- Accurately measures down to 10's of milliohms/square without external meter
- 4mm socket facility to connect an external meter
- RS232/USB connectivity for control and for collecting data in CSV format
- Automatic compliance voltage limit protects your sample which reduces maximum voltage as higher currents are set 5-40V
- Warning messages when compliance limit is reached, or DVM input exceeded
- Self test facility
- Flash upgradeable software
- CE Marked
- AD/DA converters and amplifiers for the DVM and setting the Current source and DVM, are Burr Brown 110V or 240V operation (selectable)
|
A complete four point probing system consists of the RM3-AR Test Unit shown above, combined with one of the mechanical four point probe units:
Multiposition Wafer Probe
Multi Height Probe
Microposition Probe
Multi Height Microposition Probe
Universal Probe
Hand Applied Probe
Jandel Four Point Probe System Configurations:
|
|